The Industry Choice for Affordable, High Quality Equipment, Supplies and Service

Home >> Small Equipment >> Vital Signs Monitors >> DRE Echo CO2 Monitor - Capnography

DRE Echo CO2 Monitor - Capnography

DRE Echo CO2 Monitor - Capnography

Product Information:

Call For Availability

Free Shipping on equipment orders $2,000 or more, shipped within the Continental United States. Excludes Cabinetry. Not valid with other discounts. Other restrictions may apply.
(Discount shown during checkout)


Product Description:

Effective Capnography Monitoring ( EtCO2 )

The DRE Echo CO2 provides cost-effective SPO2 and EtCO2 monitoring with accuracy and durability. The Echo CO2 ensures effective capnography monitoring for intubated and non-intubated patients for continuous long-term monitoring. It is tailored for mechanically ventilated and non-intubated patients.

Capnography is outpacing Pulse Oximetry as the new standard of care in outpatient surgery procedures.

Features:

  • 5.7 inch high resolution display for easy reading
  • Lightweight, portable design and user-friendly interface for easy operation
  • Flexible configurations to meet different clinical needs
  • Standard SpO2
  • PR measurement
  • Respironics Loflo sidestream or CAPNOSTAT 5 EtCO2 mainstream measurement
  • Powerful storage capacity
  • Real-time parameters measurement display with trend table for easy reviewing
  • Built-in rechargeable Lithium-ion battery for 10 hours continuous working
  • Bi-directional communications with central station by wired or wireless network
  • Nurse call
  • Suitable for adult, pediatric and neonatal patient

Specifications: (General)

General Size: 200.8mm (L) × 241mm (H) × 189 mm (D)
Weight: 3 kg
Display: 5.7 inches
Color TFT resolution: 640X480
Power Supply: 100-240 VAC, 50/60HZ
Pmax=70VA FUSE T 1.6AL
Battery Type: Lithium-ion
Voltage: 14.8 V DC Capacitance: 4,400 mAh
Working period: Color TFT 480min
Rechargeable period: < 360min
Thermal Recorder: (Optional)
Record Width: 48 mm
Paper Speed: 25mm/s

SpO2 (Standard)

General Measuring Range: 0 ~ 100 %
Alarm Range: 0 ~ 100 %
Resolution: 1 %
Accuracy Adult (including Pediatric):
  • ± 2% (70%~100% SpO2)
  • Undefined 0~70% SpO2
Neonate:
  • ±3% (70%~100% SpO2)
  • Undefined: (0~70% SpO2)
Pulse Rate:
  • Measuring and Alarm Range: 20~300 bpm
  • Resolution: 1 bpm
  • Accuracy: 3 bpm

Respironics CO2

Mainstream CO2 Measurement Range:
  • 0 to 150 mm Hg,0 to 19.7%,
  • 0 to 20 kPa (at 760 mm Hg)
Rise Time:
  • Less than 60 ms Adult/Infant
  • Reusable or Single Patient Use
  • Airway Adapter
CO2 Resolution:
  • 0.1 mm Hg — 0 to 69 mm Hg
  • 0.25 mm Hg — 70 to 150 mm Hg
CO2 Accuracy:
  • 0-40 mm Hg — ±2 mmHg
  • 41 - 70 mm Hg — ± 5% of reading
  • 71 - 100 mm Hg — ± 8% of reading
  • 101 - 150 mm Hg — ±10% of reading
  • Above 80 bpm — ±12% of reading
CO2 Stability:
  • Short Term Drift: Drift over four hours shall not exceed 0.8 mm Hg max.
  • Long Term Drift: Accuracy specification will be maintained over a 120-hour period

Respiration Rate Accuracy: ±1 breath

Calibration: No routine user calibration required. An airway adapter zero is required when changing to a different style of airway adapter

Water Resistance: IPX4-Splash-proof (sensor head only)

Shock Impact:
  • EN60068-2-6 Sinusoidal Vibration
  • EN60068-2-27 Shock
  • EN60068-2-64 Random Vibration
  • Able to withstand repeated 6 foot drops onto tiled floor while operating

Sidestream

Mainstream

Sample Rate: 50 m/L per minute

CO2 Measurement Range:
  • 0 - 150 mm Hg, 0 to 20 kPa (at 760 mm Hg).
  • Barometric Pressure supplied by Host
CO2 Resolution:
  • 0.1 mm Hg 0 to 69 mm Hg
  • 0.25 mm Hg 70 to 150 mm Hg
CO2 Accuracy:
  • 0 - 40 mm Hg ±2 mm Hg
  • 41 - 70 mm Hg ±5% of reading
  • 71 - 100 mm Hg ±8% of reading
  • 101 - 150 mm Hg ±10% of reading
  • Above 80 bpm ±12% of reading
CO2 Stability:
  • Short Term Drift: Drift over four hours shall not exceed 0.8 mm Hg max.
  • Long Term Drift: Accuracy specification will be maintained over a 120-hour period

Respiratory Rate Range: 2 to 150 bpm

Respiratory Rate Accuracy: ±1 breath

Sample Cell/Filter: Proprietary single patient use sample cell and inline filter are integrated with the sample line which eliminates contamination of the internal system.

Nasal Sampling Kits for Non-intubated Patients:
  • Adult, pediatric and infant nasal CO2 sampling, nasal CO2 sampling and O2 delivery
  • Adult and pediatric nasal/oral CO2sampling, nasal/oral CO2 sampling and O2 delivery
On-Airway Adapter KITS for Intubated Patients:
  • Adult/Pediatric with and without dehumidification tubing
  • Pediatric/Infant, low dead space, with and without dehumidification tubing
  • Taper meets ISO 5356-1
Sample Kit Hours of Use:
  • Nasal Cannula (all styles) — up to 12 hours
  • On-Airway Adapter Kits without dehumidification tubing — up to 12 hours

Sample Cell Detection: Insertion automatically turns sampling pump on. Removal automatically turns sampling pump off.

Water Resistance IPX4: Splash-proof (when sample cell is inserted in sample cell receptacle)

Shock ImpactIEC TR 60721-4-7 Class 7M3 (designed to withstand environments subject to significant vibrations or high shock levels)
  • EN60068-2-27 Shock
  • EN60068-2-64 Random vibration